For cleanroom validators and quality engineers who need to verify theoretical compliance thresholds before deploying air sampling probes or drafting validation protocols.
This calculator applies the ISO 14644-1:2015 standard formula Cn = 10^N * (0.1/D)^2.08 to determine maximum allowed particles per cubic meter for specific sizes.
Maximum Particle Concentration Limit:
0 particles/m³
Formula: Cn = 10N × (0.1 / D)2.08
Preliminary estimation only, not a substitute for review by a qualified/licensed engineer.
It applies the ISO 14644-1:2015 standard equation Cn = 10^N × (0.1/D)^2.08, where N is the class number and D is the particle size in µm.
No, it provides preliminary theoretical limits only and must be verified alongside your calibrated particulate counters and facility-specific SOPs.
Lower classes allow fewer particles; Class 6 reduces the permissible concentration roughly tenfold compared to Class 7 for any given particle size.