Mean Recovery Uncertainty Calculator
This calculator turns the validation summaries of up to six reference materials into recoveries. For each certified reference material or proficiency test result you enter the reference value C, its expanded uncertainty U with coverage factor k, your measured mean c, the precision standard deviation s and the number of replicates n. The page returns each recovery r = c/C, the term t that each material adds to the uncertainty, the mean recovery R (also as a percentage) and its standard uncertainty u(R).
The method follows Eurachem/CITAC MUPR 2026 (first edition), section 8.2.4, Equations 8 and 9 and Example 2 on page 13. Nothing is rounded before display. The rows are treated as independent, the reference uncertainties should be small against the target uncertainty, and the result is an uncertainty only, not a pass or fail decision.
Calculator
All concentrations are in mg/L. A row is used when its reference value C is filled in; a used row needs all six fields. Leave unused rows empty.
Results
| Material | Recovery r | Term t |
|---|---|---|
| Results appear here after you press Calculate. | ||
What the result assumes
- The rows (materials) are assumed independent of each other.
- The uncertainty of each reference value should be small against the target uncertainty of the method.
- For n below 10, s needs a validated precision model; the page uses s as entered.
- The page gives an uncertainty only, not a pass or fail decision.
How the calculation works
For each used row i the page computes the recovery and the row term, then combines the N used rows:
| Symbol | Meaning | Unit |
|---|---|---|
| C | reference value of the material (certified value or proficiency test value) | mg/L |
| U | expanded uncertainty of C | mg/L |
| k | coverage factor of the certificate; U / k is the standard uncertainty of C | dimensionless |
| c | mean concentration measured by the laboratory | mg/L |
| s | precision standard deviation | mg/L |
| n | number of replicates behind c | count |
| r | recovery of one material | dimensionless |
| t | uncertainty term of one material | dimensionless |
| N | number of used rows | count |
| R | mean recovery (100·R in percent) | dimensionless |
| u(R) | standard uncertainty of R | dimensionless |
The division by N in u(R) is outside the square root; √(Σ ti / N) would be wrong. The second term of t is the source formula with that term expanded, written with c2 and C4, so that it stays finite when c is 0. No value is rounded before display.
The formulas do not hold as written when the rows are not independent. For n below 10, s should come from a validated precision model, and the reference uncertainties should be small against the target uncertainty. The page uses the summaries as given and does not estimate s from raw replicates.
Source: Eurachem/CITAC MUPR 2026, first edition, section 8.2.4, Equations 8 and 9, Example 2, page 13.
How to use the calculator
- For each reference material (a certified reference material or a proficiency test result), enter the reference value C in mg/L, its expanded uncertainty U in mg/L and the coverage factor k stated with it.
- Enter the mean c that your method measured for that material and the precision standard deviation s, both in mg/L, and the number of replicates n.
- Use one row per material, up to six. A row counts only when its C field is filled in; leave unused rows completely empty.
- Press Calculate. If a message appears under a field, it names the row concerned; correct the value and press Calculate again.
- Read r and t for each row, then the mean recovery R, 100·R in percent and the standard uncertainty u(R), and record them with your validation documents.
Frequently asked questions
Why is u(R) divided by N outside the square root?
The mean recovery R is the sum of N recoveries divided by N, and the rows are assumed independent. Their terms t therefore add, and the division by N is applied to the square root of that sum: u(R) = sqrt(sum of t) / N. Writing sqrt(sum of t / N) instead is wrong: it gives a value larger by a factor equal to the square root of N.
Why does the calculator ask for the coverage factor k?
A certificate or proficiency test report states an expanded uncertainty U for the reference value C together with a coverage factor k. The calculation uses the standard uncertainty of C, which is U divided by k. With U = 0.1 mg/L and k = 1 the standard uncertainty is 0.1 mg/L; with k = 2 it would be 0.05 mg/L.
What if a material was measured fewer than 10 times?
For n below 10, the precision standard deviation s needs a validated precision model rather than the scatter of the few replicates alone. The page uses the summaries exactly as entered and does not estimate s from raw replicates, so enter the s you have justified for that material.
Does the result tell me whether the recovery is acceptable?
No. The page gives the mean recovery R and its standard uncertainty u(R) only, not a pass or fail. Keep in mind that the uncertainty of the reference values should be small against the target uncertainty of the method.
Disclaimer
The results are an estimate for documentation only. They depend entirely on the summaries you enter and on the assumptions stated above, and they do not replace the review of a qualified person or the requirements of your quality system.