NaOH Standardization Uncertainty Calculator
This calculator gives the concentration of a sodium hydroxide (NaOH) solution standardized against potassium hydrogen phthalate (KHP), together with its combined standard uncertainty and its expanded uncertainty. The model follows Eurachem/CITAC QUAM:2012.P1 Example A2 (pp. 41-50, Appendix E.2) and is intended for analytical laboratory staff documenting a standardization.
Five quantities contribute to the result: the repeatability of the whole procedure, the mass of KHP weighed by difference, the purity of the KHP, its molar mass and the delivered NaOH volume. Each is turned into a standard uncertainty from your laboratory data, and the combined uncertainty is obtained by shifting each quantity numerically and combining the changes in concentration.
Enter your laboratory data
Results
Mass of KHP m: — g
Standard uncertainties of the inputs
| Quantity x | Value | u(x) | Unit |
|---|---|---|---|
| Repeatability factor rep | — | — | relative (dimensionless) |
| Mass of KHP m | — | — | g |
| Purity of KHP P | — | — | mass fraction |
| Molar mass of KHP M | — | — | g/mol |
| Delivered NaOH volume V | — | — | mL |
Diagnostic only (not the reported result): first-order combined uncertainty uc = — mol/L
- The rounding stages follow the source example and can hide tiny contributions, which may then appear as zero.
- The KHP molar mass uses historical QUAM atomic weights, kept fixed.
- This page gives an uncertainty model only, not a pass or fail decision.
Calculation model
The concentration of the NaOH solution is
- cNaOH
- concentration of the NaOH solution, mol/L
- rep
- repeatability factor, equal to 1, dimensionless
- m
- mass of KHP, gross mass minus tare mass, g
- P
- purity of KHP, mass fraction
- M
- molar mass of KHP, 8 · 12.0107 + 5 · 1.00794 + 4 · 15.9994 + 39.0983 = 204.2212 g/mol
- V
- delivered NaOH volume, mL
Standard uncertainties
- Repeatability: u(rep) = RSD / 100.
- Mass: a = linearity half-width / √3, rounded to 0.01 mg; u(m) = √2 · a, rounded to 0.01 mg, then converted to g (÷ 1000).
- Purity: u(P) = certificate half-width / √3, rounded to 5 decimals.
- Molar mass: u(M) = 0.0038 g/mol, fixed.
- Volume: b = burette half-width / √6, rounded to 0.001 mL; d = Vt · 2.1 × 10−4 °C−1 · ΔT / 1.96, rounded to 0.001 mL; u(V) = √(b2 + d2), rounded to 0.001 mL. Here Vt is the volume for the temperature effect (mL) and ΔT the temperature half-width at 95 % (°C).
Combination
Each quantity xi is shifted alone by +u(xi), and the signed change Δci = c(xi + u(xi)) − c is kept. Then
The diagnostic first-order value is √Σ (c · u(xi) / xi)2. The concentration is shown to 5 decimals; uc and U are shown to 2 significant digits.
How to use the calculator
- Enter the gross mass of the container plus KHP and the tare mass of the container, both in grams. The KHP mass is their difference.
- Enter the balance linearity half-width in milligrams.
- Enter the KHP purity as a mass fraction (1 for 100 %) and the half-width from the purity certificate as a fraction.
- Enter the delivered NaOH volume, the burette calibration half-width and the volume for the temperature effect, all in millilitres, and the temperature half-width at 95 percent in degrees Celsius.
- Enter the repeatability RSD of the whole procedure in percent.
- Press Calculate and read the KHP mass, the table of standard uncertainties, the concentration, the combined standard uncertainty and the expanded uncertainty (k = 2).
Frequently asked questions
How is the combined standard uncertainty calculated?
Each of the five input quantities (repeatability, KHP mass, KHP purity, KHP molar mass and NaOH volume) is shifted alone by its standard uncertainty, and the signed change in the NaOH concentration is kept. The combined standard uncertainty is the square root of the sum of the five squared changes, and the expanded uncertainty is twice that value (coverage factor k = 2), following Eurachem/CITAC QUAM:2012.P1 Example A2.
Why are some intermediate uncertainties rounded?
The rounding stages follow the source example: the balance linearity term and the mass uncertainty are rounded to 0.01 mg, the purity uncertainty to 5 decimals, and the burette, temperature and volume uncertainties to 0.001 mL. This reproduces the published numbers, but it can hide very small contributions, which may then show as zero.
Which molar mass of KHP is used?
The molar mass is fixed at 204.2212 g/mol, calculated from the historical atomic weights used in QUAM (C 12.0107, H 1.00794, O 15.9994, K 39.0983), with a fixed standard uncertainty of 0.0038 g/mol. These values are kept so that the results match the source example.
Does the calculator tell me whether my standardization passes?
No. The page gives an uncertainty model only, not a pass or fail decision. The first-order combined uncertainty is shown as a diagnostic only, to compare with the main result.
Disclaimer: the results are an estimate for documentation only. They reproduce the uncertainty model of Eurachem/CITAC QUAM:2012.P1 Example A2 with its rounding stages and historical atomic weights, and do not replace your laboratory's validated procedures or a pass or fail decision.